About Advanced Instruments, Inc. All Advanced Products Contact Information Product Support New and Events Search our site
Fiske 110 Osmometer Instrument Specifications
Electrical:
Voltage: 100 to 130 VAC or 200 to 250 VAC
Frequency: 50 or 60 Hz
Fuse: 1.25-Amp time delay for 100-130 VAC,
1-Amp time delay for 200-250 VAC
Power: 100 watts
Memory Backup: integral lithium cell: 2 years minimum,
10 years maximum
Sample Size: 15 µL
Sample Capacity: single sample
Readout: 20-character digital display
Units: mOsm/kg H2O
Range: 0 to 2000 mOsm/kg H2O
Linearity: ±2mOsm/kg H2O between 0 and 400 mOsm/kg H2O; <0.5% between 400 and 1000 mOsm/kg H2O; <1% between 1000
and 2000 mOsm/kg H2O
Resolution: 1 mOsm/kg H2O
Repeatability: ±2 mOsm/kg H2O (1 S.D.) between 0 and 400 mOsm/kg H2O;
ñ0.5% (1 S.D.) between 400 and 1500 mOsm/kg H2O; ±1%
(1 S.D.) between 1500 and 2000 mOsm/kg H2O
Communications: RS-232C and parallel printer ports
Storage Temperature: -40 to +160°F or-40 to +70°C
Operating Temperature
(ambient):
64 to 95°F or18 to 35°C
Drift: less than 1 digit per month and 1 digit for every 9°F or 5°C
ambient temperature change
Room Humidity: 5% to 80% R.H. (non-condensing)
Start-up Time: 30 minutes from power on
Test Time: 80 seconds per sample
Dimensions:
inches
centimeters
Width:
Depth:
Height:
13
17
8
32
43
20
Weight:
pounds
kilograms
Net:
Shipping:
18
24
8
11
Warranty: One-year limited warranty on workmanship and all parts except glass, plastic and parts warranted by their makers
Certification: