| Electrical Voltage: 100 to 130 VAC or 200 to 250 VAC Frequency: 50 or 60 Hz Fuses (2): 250 VAC time delay (T): 2-Amp for 100-130 VAC, 1-Amp for 200-250 VAC Power: 150 watts Memory Backup: integral lithium cell: 10 years typical |
Storage
Temperature: -40°F to +158°F or -40°C to +70°C |
||
| Operating
Temperature (ambient): 64°F to 95°F or 18°C to 35°C |
|||
| Sample
Size: 0.2 or 0.25 mL |
Drift:
less than 1 digit per month plus 1 digit for every 9°F or 5°C ambient temperature change |
||
| Sample Capacity: single sample | |||
| Readout: 20-character digital display | Room Humidity: 5-80% (non-condensing) | ||
| Units: mOsm/kg H2O | Time: immediately, from stand-by; under 5 min- utes, from power on | ||
| Ranges Low: 0 to 2000 mOsm/kg H2O High: 1400 to 4000 mOsm/kg H2O |
Test Time: approximately 2 minutes per sample in 0 to 2000 mOsm range; moderately longer in high range | ||
| Linearity: less than ±0.5% from straight-line | Dimensions |
inches
|
centimeters
|
| Resolution: 1 mOsm/kg | Width Depth Height With head up |
13
17 12 16 |
32
43 29 41 |
| Repeatability:
0 to 400 mOsm ±2 mOsm/kg (1 S.D.) 400 to 4000 ±0.5% (1 S.D.) |
|||
| Communications:
RS-232C and parallel printer ports |
Weight |
pounds |
kilograms |
| Warranty: One-year limited warranty on workmanship and all parts except glass, plastic and parts warranted by their makers | Net Shipping |
33
38 |
15
17 |
| Certifications:
|
|||